Accurate, Cost-Effective Chemical Analysis and Monitoring

Advanced Chemical Sensors, Inc.
Technical Information Sheet

Chloroform Vapor Monitor
Catalog No.OV-30

$39.00/each

ACS Vapor Monitor can be worn as a badge near the breathing zone to measure personal exposure, or placed in a room to measure area concentration.

Occupational safety standards are based on personal exposure monitoring of each employee who may be exposed to the vapor as a regular part of their job. Periodic monitoring of personnel and room concentrations is recommended


Exposure Limits: ACGIH OSHA NIOSH
TWA (ppm) 10 N/A N/A
STEL/CEIL (ppm) N/A 50 2 (1hr)

Sampling Medium: Activated High-Purity Carbon Molecular Sieve
Method of Analysis: Gas Chromatography
Reference Method: US OSHA Method 5
Recommended Sampling Time: 15 minutes to 8 hours
Minimum Level of Quantitation: 0.2 ppm for 8 hours
7 ppm for 15 minutes
Interferences: No interferences are known.
Effect of Temperature: Less than 10% for each 10 °C variation from 24 °C
Effect of Humidity: No effects detected at 30% RH to 92% RH.
Overall System Accuracy: 11% at 10 ppm for 8 hours
16% as 50 ppm for 15 minutes
Storage Conditions: Maximum shelf life one year.
Recommended Holding Time: Two weeks prior to analysis.